Karakterisasi nosel dengan variasi sudut lubang baffle dan jarak tembak terhadap gaya impak untuk aplikasi gene gun = characterisation of nozzle through baffle holes angle and shooting distance variations to impact force for application of gene gun

(Fakultas Teknik Universitas Indonesia, 2012)

 Abstrak

Pengembangan gene gun sebagai instrumen administrasi vaksin
intradermal memerlukan banyak penelitian di berbagai hal, salah satunya pada
pengaplikasian teknologi nosel terhadap dampak yang ditimbulkan. Pengujian
dilakukan untuk mendapatkan desain nosel yang memiliki keluaran gaya impak
terendah dengan mengetahui karakteristik nosel melalui pengaruh sudut lubang
baffle dan jarak tembak terhadap gaya impak. Variasi nosel yang diuji adalah
nosel standar tanpa lubang baffle dan nosel dengan sudut lubang baffle sebesar
30o, 90o, dan 150o. Keempat nosel tersebut diuji pada tiga jarak tembak, yaitu
0 mm, 10 mm, dan 20 mm. Hasil eksperimen dan analisa secara statistik
menunjukkan pola yang menyatakan adanya hubungan antara variasi sudut lubang
baffle dan jarak tembak terhadap gaya impak dengan adanya indikasi titik jarak
tembak optimal untuk tiap variasi nosel dan kecenderungan menurunnya gaya
impak seiring dengan meningkatnya besar sudut lubang baffle. Adapun besar gaya
impak terkecil dihasilkan oleh nosel dengan sudut lubang baffle 150o pada jarak
tembak 0 mm. Metode simulasi komputasi memberikan hasil berupa profil
kecepatan dan nilai dari beberapa parameter aliran di dalam dan luar nosel yang
menunjukkan bahwa nosel yang digunakan termasuk kategori nosel supersonik.

Abstract
There have been many researches in demand to develop gene gun as an
intradermal vaccine administration instrument. One of those researches is in the
application of nozzle technology with its implication. Tests to investigate nozzle
characteristics in terms of baffle holes angle and shooting distance variation to
impact force have been carried out. Nozzles under test were standard nozzle
without baffle holes and nozzle with baffle holes angle 30o, 90 o, and 150 o. These
four nozzles were tested at three shooting distances, which were 0 mm, 10 mm
and 20 mm. Patterns that stated relationship between the baffle holes angle and
shooting distance that generate the impact force have been found through
experiments and statistical analysis. They indicated the existence of optimum
shooting distance for each nozzles. In addition, the impact force trend is
decreasing when the baffle holes angle is widened. It is also discovered that the
lowest impact force was generated by nozzle with baffle holes angle 150o with
shooting distance 0 mm. Computational simulations were conducted to investigate
the velocity profile and some fluid flow parameter values inside and outside the
nozzles. From these, it can be concluded that the nozzles used for the application
of gene gun as intradermal vaccine administration instrument should be
considered as supersonic nozzle.

 File Digital: 1

 Metadata

Jenis Koleksi : UI - Skripsi Open
No. Panggil : S43508
Program Studi :
Penerbitan : [Place of publication not identified]: Fakultas Teknik Universitas Indonesia, 2012
Bahasa : ind
Sumber Pengatalogan :
Tipe Konten :
Tipe Media :
Tipe Carrier :
Deskripsi Fisik : xiv, 80 hlm. : ill. ; 28 cm + lamp.
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S43508 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20324021
Cover