Verifikasi akurasi posisi pasien IMRT Kanker Nasofaring dengan registrasi citra DRR/EPID = The patient positioning accuracy verification of IMRT technique of Nasopharyngeal Cancer with DRR/EPID image registration

Susi Nofridianita; Supriyanto Ardjo Pawiro, supervisor (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2013)

 Abstrak

Radioterapi merupakan salah satu modalitas penanganan Kanker Nasofaring. Kemajuan teknologi radioterapi termasuk perkembangan intensity modulated radiotherapy (IMRT) memberikan hasil yang cukup memuaskan dalam penanganan Kanker Nasofaring baik secara klinis maupun dosimetri. IMRT dapat mengurangi efek akut dan kronik, dengan cakupan dosis maksimal pada tumor dan dosis minimal pada organ/jaringan sehat di sekitarnya.
Tujuan dari penelitian ini adalah untuk mengetahui perbandingan verifikasi akurasi posisi pasien IMRT Kanker Nasofaring dengan registrasi citra DRR/EPID. Analisis data retrospektif terhadap citra DRR/EPID dari 35 pasien proyeksi AP dan Lateral (140 citra) yang telah diverifikasi secara manual kemudian secara simulatif diterapkan metode fusi semiotomatis dengan program FIJI. Penggunaan program FIJI dapat memperbaiki kualitas citra DRR/EPID sehingga memudahkan dalam verifikasi geometri radioterapi.
Hasil penelitian ini secara statistik menunjukkan tidak terdapat perbedaan antara verifikasi manual dengan metode fusi pada radioterapi pasien kanker nasofaring, namun secara deskriptif terdapat kecenderungan bahwa metode fusi denganprogram FIJI memberikan verifikasi geometri radioterapi yang lebih baik dibandingkan metode manual. Verifikasi akurasi posisi pasien sangat penting dilakukan karena besarnya translasi sangat mempengaruhi PTV dan dosis yang diterima jaringan sehat disekitarnya serta organ beresiko.

Radiotherapy is one of common treatment modality for Nasopharyngeal Cancer. The development of intensity modulated radiotherapy (IMRT technique) gives satisfactory results in the nasopharyngeal cancer treatment, both clinically and dosimetry. IMRT can reduce the effects of acute and chronic, with a maximum dose coverage to the tumor and minimal dose to the organ or normal tissue surrounding target value.
The purpose of this study is to compare the accuracy of patient positioning verification of Nasopharyngeal Cancer IMRT with DRR / EPID image registration. Retrospective data analysis of the AP and Lateral projections DRR and EPID images 35 patients (140 images) were then manually verified by simulative applied fusion semiautomatic with FIJI program. FIJI program improved the image quality of the DRR and EPID to facilitate the image registration.
Results of this study shows no statistically significant difference between the manual verification and semiautomatic fusion method of nasopharyngeal cancer patients, but there is a tendency that the semiautomatic method with FIJI program provides verification geometry radiotherapy better a result than manual methods. Accuracy verification of patient positioning is very important because it greatly affects the magnitude of translational PTV and the dose received by surrounding healthy tissue and organs at risk.

 File Digital: 1

Shelf
 S47598-susi nofridianita.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Skripsi Membership
No. Panggil : S47598
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2013
Bahasa : ind
Sumber Pengatalogan :
Tipe Konten :
Tipe Media :
Tipe Carrier :
Deskripsi Fisik : xiii, 51 hlm. : il. ; 28 cm. + lamp.
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S47598 14-23-29665232 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20345547
Cover