UI - Tesis Membership :: Kembali

UI - Tesis Membership :: Kembali

Optimalisasi kejelasan anatomi dan dosis radiasi pada CT Scan tulang wajah (maxillofacial) dengan variasi kuat arus tabung = Optimization of anatomy clarity and radiation dose at face bone scanning (maxillofacial) by variation of tube current

Agustina Dwi Prastanti; Cholid Badri, supervisor; Jeffri Ardiyanto, supervisor; Anwar S. Ibrahim, examiner; Boy Subirosa Sabarguna; Supardjo, examiner (Fakultas Kedokteran Universitas Indonesia, 2013)

 Abstrak

Tujuan penelitian ini adalah untuk menganalisis perbedaan variasi nilai kuat arus tabung terhadap kejelasan anatomi tulang wajah dan dosis radiasi. Jenis penelitian ini adalah penelitian eksperimental. Data diambil dari empat variasi penggunaan nilai arus tabung (mA) 200 mA, 150 mA, 100 mA dan 50 mA dengan parameter yang lain konstan. Dosis radiasi diukur dengan CTDI. Gambar dinilai oleh responden yang terdiri dari 20 (duapuluh) Dokter Spesialis Radiologi yang tidak menyadari tentang pengaturan kuat arus tabung pada gambar yang dihasilkan. Kualitas gambar dianalisis dengan metode skoring pada 8 (delapan) kriteria anatomi. Palatum, struktur trabekula tulang dan kortex, sinus paranasal, dinding orbita lateral dan medial, orbital roof dan orbital floor, zygomatic, nasal cavity dan ethmoid dinilai dengan skor 1 jika tidak jelas, skor 2 jika jelas dan skor 3 jika sangat jelas. Tidak ada perbedaan yang signifikan pada keempat kelompok dalam menentukan kejelasan anatomi tulang wajah berdasarkan sistem skoring yang digunakan dalam penelitian ini. Dosis radiasi dari penilaian CTDI menunjukkan bahwa dosis dapat dikurangi sebesar 75% pada penggunaan kuat arus tabung 50 mA atau 11,40 mGy dari arus protokol standar 200 mA atau 45,61 mGy. Hal ini sangat penting untuk mengurangi resiko kebutaan pada lensa mata.

The purpose of this study was to analyze the differences in the variation of tube current of the clarity of the facial bones anatomy and radiation dose. This research is an experimental study. Data were taken from four variations use the value of tube current (mA) 200 mA, 150 mA, 100 mA and 50 mA with the other parameters constant. CTDI measured radiation dose. Images assessed by respondents consisted of 20 (twenty) Radiology Specialists who are unaware of the settings on the tube current of the resulting image. The image quality was analyzed by the method of scoring in 8 (eight) anatomical criteria. Palate, structure of trabecular bone and cortex, paranasal sinuses, lateral and medial orbital wall, orbital roof and orbital floor, zygomatic, nasal cavity and ethmoid assessed with a score of 1 if it is not obvious, a score of 2 if it is clear and score 3 if very clear. There is no significant difference in the four groups in determining the clarity of the facial bones anatomy based on the scoring system used in this study. Radiation dose from CTDI assessment showed that the dose can be reduced by 75% in the use of tube current of 50 mA or 11.40 mGy of 200 mA current standard protocol or 45.61 mGy. It is very important to reduce the risk of blindness in the eye lens.

 File Digital: 1

Shelf
 T-Pdf Agustina Dwi Prastanti.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Tesis Membership
No. Panggil : T-Pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : Jakarta: Fakultas Kedokteran Universitas Indonesia, 2013
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xvi, 93 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
T-Pdf 15-19-834356179 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20350786
Cover