Evaluasi citra dan estimasi dosis pada sistem on board imager CBCT linac di MRCCC Siloam = Image evaluation and dose estimation of on board imager system CBCT linac at MRCCC Siloam

Priska Andini Putri; Supriyanto Ardjo Pawiro, supervisor; Rina Taurisia, supervisor; Djarwani Soeharso Soejoko, examiner; Sugiyantari, examiner ([, Universitas Indonesia], 2014)

 Abstrak

[ABSTRAK
Cone Beam CT adalah perangkat citra pemandu yang diintegrasikan pada perangkat LINAC radioterapi. Perangkat tersebut banyak digunakan untuk verikasi posisi pasien dalam tindakan radioterapi. Dalam penelitian ini telah dilakukan untuk estimasi dosis pada daerah kepala, dada, dan pelvis dan untuk evaluasi citra dengan menggunakan perlakuan Cone Beam CT satu putaran penuh. Fantom rando digunakan pada penelitian ini untuk mensimulasikan keadaan yang mendekati sebenarnya dengan kondisi klinis 100 kVp 145 mAs, 110 kVp 262 mAs, dan 125 kVp 680 mAs untuk berturut-turut prosedur perlakuan CBCT kepala, dada, dan pelvis. Dosimeter yang digunakan dalam penelitian ini adalah TLD yang dikalibrasi di PTKMR BATAN dengan kondisi yang sesuai dengan kondisi klinisnya. Fantom Catphan 504 digunakan untuk mengevaluasi citra CBCT dengan menggunakan kondisi klinis protokol pelvis, yaitu dengan energi 125 kV. Hasil estimasi dosis yang diperoleh dari penelitian ini adalah 4.018±0.334 mGy, 4.210±0.428 mGy, dan 12.547±3.046 mGy berturut-turut pada kepala, dada, dan pelvis. Hasil citra dari penelitian ini dievaluasi menggunakan Image J yang menghasilkan nilai densitas material yang ada pada fantom Catphan mendekati nilai acuannya, didapatkan ketebalan slice sebesar 2.153 mm, resolusi uji citranya sebesar 5 lp/cm, dan uji uniformitas pada pusat ROI adalah 34.610±40.999 HU.

ABSTRACT
, Cone Beam CT is an image guided device which is integrated in radiotherapy
LINAC. The device often used to verify the positions of patient in radiotherapy
treatment. Dose estimation for three different clinically scan sites (head, thorax,
and pelvis) and image evaluation were performed during the research using full
circular treatment by Cone Beam CT. Rando phantom was also used in this
research to simulate the real condition for clinical scans (100 kVp 145 mAs for
head, 110 kVp 262 mAs for thorax, and 125 kVp 680 mAs for pelvis). TLDs were
used and calibrated at PTKMR BATAN with the real condition as clinical
condition to mesure the dose. For image evaluation were performed using pelvis
clinical scan (125 kV) and using Catphan 504 phantom. Results of dose
estimation are 4.018±0.334 mGy for head, 4.210±0.428 mGy for thorax, and
12.547±3.046 mGy for pelvis. Results of image evaluation which was using
Image J generated the approximate recommendation value of material density on
the Catphan Phantom; slice thickness is 2.153 mm, high resolution is 7 lp/cm, and
uniformity in center ROI is 34.610±40.999 HU]

 File Digital: 1

Shelf
 S54790-Priska Andini Putri.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Skripsi Membership
No. Panggil : S54790
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : [Place of publication not identified]: [, Universitas Indonesia], 2014
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resources
Deskripsi Fisik : xv, 41 pages : illustration ; 30 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S54790 14-18-464341466 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20385629
Cover