Pengukuran profil dosis sinar-X computed tomography dengan variasi pitch menggunakan film gafchromic XR QA2 dan thermoluminescence dosimeter (TLD) = Measurement of dose profile computed tomography X ray with pitch variation using XR QA2 gafchromic and thermoluminescence dosimeter (TLD) / Septi Purwaningsih

Septi Purwaningsih; Supriyanto Ardjo Pawiro, supervisor; Djarwani Soeharso Soejoko, co-promotor; Adhi Harmoko Saputro, examiner; I Putu Susila, examiner; Warsito, examiner ([Publisher not identified] , 2015)

 Abstrak

[ABSTRAK
Penelitian ini bertujuan untuk mengetahui profil dosis sinar-x Computed
Tomography dengan variasi pitch menggunakan film Gafchromic XR QA2 dan
TLD pada fantom kepala orang dewasa dan kepala pediatrik. Untuk tujuan
tersebut dilakukan pengukuran dan analisis profil dosis di sepanjang sumbu z
rotasi scan pada pusat fantom dan beberapa tepi di kedalaman 1 cm, analisis profil
dosis pada posisi yang sama dengan nilai pitch yang berbeda, dan analisis profil
dosis pada kepala orang dewasa dan kepala pediatrik. Dari pengukuran dan
analisis data diperoleh hasil profil dosis pada scan kepala orang dewasa dan
kepala anak-anak dengan nilai dosis maksimal di tengah, menurun pelan ke arah
tepi dengan tendensi cenderung simetri. Profil dosis dengan variasi nilai pitch,
diperoleh nilai dosis yang semakin menurun dengan bertambahnya nilai pitch.
Profil dosis pada scan kepala orang dewasa memiliki nilai dosis lebih besar dari
pada dosis pada scan kepala pediatrik, karena scan kepala orang dewasa
menggunakan mAs lebih besar. Hasil pengukuran menggunakan TLD diperoleh
data yang bersifat disktrit, namun memiliki pola distribusi dan nilai dosis yang
hampir sama dengan hasil dari pengukuran menggunakan film Gafchromic. Hasil
pengukuran profil dosis menggunakan film Gafchromic didapatkan grafik
kontinu. Nilai dosis maksimum hasil scan didapatkan pada posisi jam 12 dan
minimum pada posisi jam 6. Profil dosis pada pitch 0,75; 1 dan 1,5 mengalami
fluktuasi dosis dengan amplitudo pada pitch 0,75 dan 1,5 mempunyai nilai yang
lebih besar jika dibandingkan dengan amplitudo pada pitch 1.

ABSTRACT
This research aims to check pattern dose profile on the adult and pediatric
head scan. Comparing result measurement dose profile along the z- axis rotation
at a depth 1 cm and center phantom with variety pitch. Measurement using
cylinder PMMA homogeneous phantom diameter 16 cm and 10 cm using XR
QA2 Gafchromic film and TLD. Result of research get dose profile adult and
pediatric head scan have same pattern, the maximum dose in the middle and
tendency symmetry in the edge. Value dose proportional with value of pitch. Dose
in the adult head more than dose pediatric head because dose in adult head scan
using bigger mAs. Dose of TLD measurement values obtained are discrete, but
has a distribution and dose values are almost the same as the measurement results
in XR QA2 Gafchromic film. Result of Gafchromic film measurement is continue
graphic. The maximum dose of the scan results at the 12 o'clock position and
minimum at 6 o'clock position. Result of research get fluctuated dose and
amplitude of pitch 0.75 and 1.5 have greater than amplitude pitch 1., This research aims to check pattern dose profile on the adult and pediatric
head scan. Comparing result measurement dose profile along the z- axis rotation
at a depth 1 cm and center phantom with variety pitch. Measurement using
cylinder PMMA homogeneous phantom diameter 16 cm and 10 cm using XR
QA2 Gafchromic film and TLD. Result of research get dose profile adult and
pediatric head scan have same pattern, the maximum dose in the middle and
tendency symmetry in the edge. Value dose proportional with value of pitch. Dose
in the adult head more than dose pediatric head because dose in adult head scan
using bigger mAs. Dose of TLD measurement values obtained are discrete, but
has a distribution and dose values are almost the same as the measurement results
in XR QA2 Gafchromic film. Result of Gafchromic film measurement is continue
graphic. The maximum dose of the scan results at the 12 o'clock position and
minimum at 6 o'clock position. Result of research get fluctuated dose and
amplitude of pitch 0.75 and 1.5 have greater than amplitude pitch 1.]

 File Digital: 1

Shelf
 T43843-Septi Purwaningsih.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Tesis Membership
No. Panggil : T43843
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : [Place of publication not identified]: [Publisher not identified], 2015
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : xix, 121 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
T43843 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20414565
Cover