Penentuan karakteristik berkas foton lapangan kecil dengan penyinaran setengah lapangan half-beam radioterapi = Determination of small field photon beam characteristic on half beam radiotherapy

Risayanti; Supriyanto Ardjo Pawiro, supervisor; Wahyu Edy Wibowo, supervisor; Djarwani Soeharso Soejoko, examiner; Nainggolan, Andreas, examiner ([Publisher not identified] , 2016)

 Abstrak

ABSTRACT
Telah dilakukan pengukuran PDD, faktor keluaran dan profil dosis pada lapangan kecil dengan teknik penyinaran setengah lapangan menggunakan Film Gafchromic EBT3, bilik ionisasi Exradin A16, dan detektor matriks PTW 2D-Array 729. Tujuan dari penelitian ini adalah mengetahui karakteristik berkas foton lapangan kecil menggunakan teknik penyinaran setengah lapangan, dan membandingkan hasil pengukuran dengan data referensi yang menggunakan teknik penyinaran lapangan penuh full-beam . Evaluasi pengukuran PDD dilakukan dengan menentukan nilai dmax dan menghitung nilai TPR20,10. Profil dosis dianalisa berdasarkan nilai FWHM dan penumbra. Hasil pengukuran menunjukkan bahwa nilai dmax, TPR20,10, FWHM, penumbra, dan faktor keluaran menghasilkan pola yang sama dengan hasil pengukuran teknik penyinaran lapangan penuh full-beam . Deviasi FWHM terbesar menggunakana Film Gafchromic EBT3 dan Exradin A16 adalah -10.99 dan -33.14 terjadi saat pengukuran pada luas lapangan 0.8 0.8 cm2 dan kedalaman 1.5 cm, sedangkan deviasi FWHM terbesar menggunakan detektor matriks PTW 2D-Array 729 adalah -46.25 terjadi saat pengukuran pada luas lapangan 1.6 1.6 cm2 dan kedalaman 1.5 cm. Persentase deviasi faktor keluaran terbesar adalah -54.22 yaitu saat pengukuran menggunakan detektor matriks PTW 2D-Array 729 pada kondisi pengukuran luas lapangan 0.8 0.8 cm2 dan kedalaman 10 cm.

ABSTRACT
PDD measurements, output factor, and dose profile have been conducted in small field with half beam irradiation technique using Gafchromic EBT3 Film, Exradin A16 ionization chamber, and PTW 2D Array 729 matrix detector. This study was aimed to determine the characteristic of small field photon beam using half beam irradiation technique and compare the measurement results with reference data that performed full beam irradiation technique. PDD measurements evaluation has been done by determining the value of dmax and calculate the TPR20,10 value. Dose profile was analyzed based on the value of FWHM and penumbra. The measurement results showed that the value of dmax, TPR20,10, FWHM, penumbra, and output factor have similar pattern with the measurement results of full beam irradiation technique. The highest FWHM deviation value that can be obtained using Gafchromic EBT3 Film and Exradin A16 are 10.99 and 33.14 occurred when measurements were conducted in the field size of 0.8 0.8 cm2 and depth of 1.5 cm. The highest deviation percentage of output factor that can be obtained using PTW 2D Array 729 matrix detector is 54.22 occured in the field size of 0.8 0.8 cm2 and depth of 10 cm.

 File Digital: 1

Shelf
 S66624-Risayanti.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Skripsi Membership
No. Panggil : S66624
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : [Place of publication not identified]: [Publisher not identified], 2016
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xvii, 86 pages : illustration ; 30 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S66624 14-18-558979411 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20444362
Cover