Estimasi dosis radiasi kepala dan abdomen pada prosedur angiografi rotasi 3 dimensi = The estimation of head and abdomen radiation dose in three dimensional rational angiography

Hamdi Sahal; Lubis, Lukmanda Evan, supervisor; Djarwani Soeharso Soejoko, examiner; Supriyanto Ardjo Pawiro, examiner; Kardinah, examiner (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2017)

 Abstrak

Pesawat angiografi rotasi 3 dimensi digunakan dalam radiologi intervensi, kardiologi intervensi dan bedah invasif minimal yang dapat memvisualisasikan pembuluh darah, dan mengevaluasi anatomi tubuh yang lebih rumit dengan dosis radiasi yang lebih besar dibandingkan generasi sebelumnya. Oleh karena itu, perlu dilakukan perhitungan estimasi dosis radiasi yang masuk ke dalam tubuh pasien yang dibandingkan dengan dosis ambang yang dapat diterima. Penelitian dilakukan dengan menggunakan phantom rando jenis wanita di dua instalasi cathlab dengan jenis pesawat yang berbeda. Pengukuran dosis dilakukan dengan menggunakan TLD pada mata, tiroid, spinal cord, payudara dan gonad pada mode preset yang berbeda untuk kepala dan abdomen.
Hasil penelitian menunjukkan pada pesawat 1, dosis yang diterima untuk pengukuran kepala pada mode 5sDR Head berkisar antara 1,17 mGy-3,68 mGy dan pada mode 5sDR Head Care berkisar antara 0,58 mGy-1,15 mGy. Sedangkan untuk pengukuran abdomen pada mode 5sDR Body dosis yang diterima adalah berkisar antara 0,50 mGy-0,85 mGy dan pada mode 5sDR Body Care berkisar antara 0,55 mGy-0,79 mGy. Pada pesawat 2, dosis yang diterima untuk pengukuran kepala pada mode Carotid Prop Scan berkisar antara 2,20 mGy-3.71 mGy dan mode Carotid Roll Scan berkisar antara 2,02 mGy-4,59 mGy. Sedangkan dosis yang diterima untuk pengukuran abdomen pada mode Abdomen Prop Scan berkisar antara 0,44 mGy-2,34 mGy dan pada mode Abdomen Roll Scan berkisar antara 0,43 mGy-1,30 mGy. Kesimpulan dari penelitian ini adalah semua mode preset tidak memberikan dosis yang mendekati dosis ambang untuk setiap titik pengukuran.

Three dimensional Rotational Angiography 3DRA is mostly used in interventional radiology, interventional cardiology, and minimal invansive surgery to visualize blood vessels, and more complicated anatomy with more visual capacity than the previous generation. The rotating nature of image acquisition with suspected high radiation dose requires dose estimation. This study was aimed to measure radiation dose in 3DRA and compare it to the thresholds for deterministic risks. Measurement using TLDs were carried out on female Rando phantom in two angiography suites with different device types, with the organ of interest being eyes, thyroid, spinal cord, breast and gonad. Different preset modes for head and abdomen were employed for comparison.
The result for device 1 showed that dose on 5sDR Head mode ranged from 1,17 mGy 3.68 mGy and in 5sDR Head Care mode ranged from 0,58 mGy 1,15 mGy while the measured dose on the body in 5sDR Body mode ranged from 0,50 mGy 0,85 mGy and in 5sDR Body Care mode ranged from 0,55 mGy 0.79 mGy. On device 2, the result showed the measured dose on the head in carotid prop scan mode ranged from 2,20 mGy 3.71 mGy and in carotid roll scan mode ranged from 2,02 mGy 4,59 mGy while the measured dose on the body in abdomen prop scan mode ranged from 0,44 mGy 2,34 mGy and in abdomen roll scan mode ranged from 0,43 mGy 1,30 mGy. The study presents that all preset modes do not deliver near threshold doses for each measurement point.

 File Digital: 1

Shelf
 S68072-Hamdi Sahal.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Skripsi Membership
No. Panggil : S68072
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2017
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xvi, 37 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S68072 14-19-266177264 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20456453
Cover