Evaluasi Dosis Neutron pada Beberapa Fasilitas Linear Accelerator Medik = Evaluation of Neutron Doses at Several Medical Linear Accelerator Facilities

Mely Mediawati; Supriyanto Ardjo Pawiro, supervisor; Djarwani Soeharso Soejoko, examiner; Dwi Seno Kuncoro Sihono, examiner; Yus Rusdian Akhmad, examiner (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2021)

 Abstrak

Tesis ini membahas dosis neutron ekuivalen pada beberapa pesawat Linear Accelerator di bawah kepala Linac dengan variasi luas lapangan dan beberapa titik pengukuran di sekitar ruang treatment. Pengukuran menggunakan dosimeter pasangan TLD 100 dan TLD 600 dengan moderator berbentuk bola dari bahan polyethylene diameter 8 inch untuk pengukuran di bawah kepala Linac. TLD disinari dengan sumber gamma 60Co untuk memperoleh faktor koreksi gamma dan dikalibrasi dengan 252Cf untuk memperoleh faktor kalibrasi neutron. Dosis neutron terbesar di bawah kepala Linac yaitu pada pesawat Linac energi 15 MV dengan luas lapangan 20 × 20 cm2 sebesar 0.35 mSv/Gy. Nilai dosis ekuivalen di pintu luar ruang treatment berkisar antara (1,37 × 10-5 – 4,23 × 10-4) μSv/MU dan di dinding operator berkisar antara (4,30 × 10-6 – 2,42 × 10-4) μSv/MU Kontribusi neutron pada dosis efektif per tahun untuk beberapa pesawat Linac di dinding operator adalah 0,1% - 2,1% dan di pintu luar ruang treatment adalah 0,3% - 2,2% dari nilai batas dosis yang ditetapkan

This thesis discusses the neutron dose equivalent on several Linear Accelerators in the isocentre with the field size variation and some measurement points around the treatment room. The measurements are performed using TLD 100 and TLD 600 dosimeter which is inserted in a ball-shaped moderator of an 8-inch diameter polyethylene material for measurements in the isocentre. TLDs are calibrated with the 60Co gamma source to find gamma correction factors and calibrated with 252Cf to find neutron calibration factors. The largest neutron dose in the isocentre was on a 15 MV energy Linac with a field size of 20 × 20 cm2 of 0.35 mSv/Gy. The equivalent dose values at the outer door of the treatment room are the range from (1.37 × 10-5 – 4.23 × 10-4) μSv/MU, whereasit is in the range from (4.30 × 10-6 – 2.42 × 10-4) μSv/MU on the operator wall. In addition neutron contribution at effective doses per year for some Linac on the operator's wall is 0,1% - 2,1%, and at the outer door of the treatment room is 0,3% - 2,2% of the set dose limit.

 File Digital: 1

Shelf
 T-Mely Mediawati.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Tesis Membership
No. Panggil : T-Pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2021
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource (rdacarries)
Deskripsi Fisik : xii, 52 pages : illustration ; appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
T-Pdf 15-22-43535864 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20514906
Cover