Optimisasi Media Kontras Pada Angiografi Jantung Dengan Parameter Signal Difference To Noise Ratio = Optimisasi Media Kontras Pada Angiografi Jantung Dengan Parameter Signal Difference To Noise Ratio

Lukman; Lubis, Lukmanda Evan, supervisor; Djarwani Soeharso Soejoko, supervisor; Supriyanto Ardjo Pawiro, examiner; Prawito, examiner; Deni Hardiansyah, examiner (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2023)

 Abstrak

Kateterisasi jantung menggunakan zat kontras untuk memandu prosedur terapeutik ke jantung dan pembuluh darah. Karena sifat agen kontras yang membebani secara biologis, penggunaan agen kontras harus dibatasi seminimal mungkin dengan hasil gambar yang dapat diidentifikasi. Sementara ambang berbasis fisiologis tersedia, ambang berbasis gambar masih belum tersedia. Oleh karena itu, perlu dilakukan analisis ambang batas volume dan debit untuk penggunaan bahan kontras. Studi pendahuluan pertama kali dilakukan untuk menentukan cairan buatan yang akan digunakan sebagai pengganti darah yang sebenarnya. Menggunakan dosing pump yang dilengkapi in-house blood flow phantom setebal 3 cm ketebalan jantung dan slab phantom setebal 23 cm ketebalan dada, mesin injektor menyuntikkan volume yang diatur dengan variasi 3, 5, 7, 10, dan 13 mL, dan variasi debit 1, 2, 3, 4, dan 5 mL/s. Hasil citra diolah menggunakan aplikasi ImageJ dengan perhitungan SDNR. Studi pendahuluan telah menunjukkan bahwa natrium klorida paling cocok untuk digunakan sebagai pengganti darah. Hasil penggunaan NaCl 0,9% sebagai darah menunjukkan bahwa ambang penggunaan zat kontras berada pada kisaran 7 mL dengan kecepatan 3 mL/s. Penggunaan agen kontras dengan volume dan laju alir yang lebih tinggi tidak menunjukkan kontras yang lebih tinggi.

Cardiac catheterization uses contrast agents to guide therapeutic procedures to the heart and blood vessels. Due to the biologically burdening nature of contrast agents, the use of contrast agents should be limited to be minimal with identifiable image results. While physiologically based thresholds are available, image-based thresholds are still unavailable. Therefore, it is necessary to analyze the threshold of volume and flowrate for the use of contrast agents. A preliminary study was first conducted to determine the artificial liquid to be used in place of actual blood. Using dosing pump-equipped in-house blood flow phantom 3 cm thick as heart thickness and slab phantom 23 cm thick as chest thickness, the injector machine injects the regulated volumes with variations of 3, 5, 7, 10, and 13 mL, and varied flowrate of 1, 2, 3, 4, and 5 mL/s. The image results were processed using the ImageJ application with SDNR calculations. Preliminary study has shown that natrium chloride was best suited for use in place of blood. Results using NaCl 0.9% as blood shown that the threshold for the use of contrast agent was in the range of 7 mL at a speed of 3 mL/s. The use of contrast agents with higher volumes and flowrates did not demonstrate higher contrast.

 File Digital: 1

Shelf
 T-Lukman.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Tesis Membership
No. Panggil : T-pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2023
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : xi, 43 pages : illustration + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
T-pdf 15-24-27339713 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 9999920536729
Cover