Sensor Kolesterol Non-Enzimatik Berbasis Nanokomposit Beta-Siklodekstrin-Sitrat/Fe3O4 serta Aplikasinya pada Sampel Kornet Sapi = Non-Enzymatic Cholesterol Sensor Based on Citrate-Modified Beta-Cyclodextrin/Fe3O4 Nanocomposite and Its Application on Corned Beef Sample

Metya Indah Firmanti; Ivandini Tribidasari Anggraningrum, supervisor; Endang Saepudin, supervisor; Tambunan, Usman Sumo Friend, examiner; Iman Abdullah, examiner (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2021)

 Abstrak

Pendeteksian kadar kolesterol, umumnya menggunakan biosensor berbasis enzim yang memerlukan biaya mahal serta perhatian khusus dalam penyimpanannya. Penelitian ini bertujuan untuk mengembangkan serta mempelajari lebih lanjut sensor elektrokimia dan spektrofotometri non-enzimatik berbasis nanokomposit β-siklodekstrin-sitrat(BCD-CIT)/Fe3O4 serta aplikasinya pada sampel kornet sapi berdasarkan prinsip pelepasan elektron pada reduksi metilen biru sebagai indikator redoks. Penelitian dimulai dengan sintesis nanokomposit yang berhasil dikarakterisasi menggunakan FTIR, XRD, SEM-EDX, dan spektrofotometri UV-Vis. Kemudian pengukuran kolesterol dilakukan pada kondisi optimum menggunakan nanokomposit BCD-CIT(3%)/Fe3O4 dengan berat 3% (w/w) dan waktu kontak 10 menit. Pengukuran menggunakan teknik amperometri dilakukan dengan screen-printed carbon electrode (SPCE). Potensial yang digunakan adalah -0,43 V dengan waktu pengukuran 90 s. Pengukuran ini menghasilkan linieritas yang baik (R2 > 0,99) pada rentang konsentrasi 0—100 μM dengan batas deteksi sebesar 3,932 μM dan batas kuantisasi sebesar 13,05 μM. Sebagai perbandingan, pengukuran kolesterol secara spektrofotometri menghasilkan linieritas yang lebih rendah pada rentang konsentrasi yang sama (0,98 < R2 < 0,99) dengan batas deteksi sebesar 15,270 μM dan batas kuantisasi sebesar 50,921 μM. Studi interferensi sensor secara komputasi pada berbagai golongan senyawa asam amino, asam lemak, dan vitamin menunjukkan bahwa beberapa senyawa yang ada dalam sampel kornet sapi dapat mengganggu kinerja sensor kolesterol. Pengukuran secara elektrokimia membuktikan bahwa kehadiran senyawa interferen memberikan perubahan respon arus yang signifikan. Aplikasi sensor dilakukan dengan pengukuran kadar kolesterol dalam sampel kornet sapi. Hasil yang diperoleh adalah 442,86 μM. Validasi menggunakan metode HPLC menunjukkan hasil yang diperoleh lebih rendah sebesar 12%.

The detections of cholesterol level generally use enzyme-based biosensors that require expensive costs as well as special attention on their storage. Therefore, this study aimed to develop electrochemical and spectrophotometric based non-enzymatic sensor using nanocomposite β-cyclodextrin-citrate(BCD-CIT)/Fe3O4 and its application on corned beef sample by detecting the electron released of reduced methylene blue as redox indicator. The research began by nanocomposites synthesis and characterization using FTIR, XRD, SEM-EDX, and UV-Visible spectrophotometry. Cholesterol measurements were done under optimum condition using 3% (w/w) nanocomposites BCD-CIT(3%)/Fe3O4 and contact time of 10 minutes. Amperometric measurements were performed using screen-printed carbon electrode (SPCE), applied potential of -0.43 V, and measurement time of 90 s. The measurement delivered good linearity (R2 > 0.99) on the concentration range of 0–100 μM. The detection limit and quantization limit are 3.932 μM and 13.050 μM, respectively. Amperometry method measurement of cholesterol used SPCE as electrode with applied potential -0.43 V for 90 s. In comparison, spectrophotometry method for cholesterol measurement resulted in lower linearity on similar concentration range (0.98 < R2 < 0.99). This method’s detection limit and quantization limit were 15.270 μM and 50.921 μM, respectively. The computational study of sensor interference was conducted on various groups of amino acids, fatty acids, and vitamins show that some compounds present in the samples may interfere the performance of cholesterol sensors. Electrochemically measurements proved that the presence of interfern compounds provided significant changes in current response. Sensor application was carried out by measuring cholesterol levels on corned beef sample. The result was 442.86 μM which showed lower yield by 12% compared to HPLC method result.

 File Digital: 1

Shelf
 S-Metya Indah Firmanti.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Skripsi Membership
No. Panggil : S-pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2021
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : x, 91 pages : illustration + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
S-pdf 14-25-12640589 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 9999920555535
Cover