Protokol Kualifikasi Instalasi (KI) dan Kualifikasi Operasional (KO) Micro Balance = Micro Balance Installation Qualification (KI) and Operational Qualification (KO) Protocols

Muhammad Mishbahus Surur; Catur Jatmika, supervisor (Fakultas Farmasi Universitas Indonesia, 2024)

 Abstrak

Proses kualifikasi peralatan merupakan langkah penting dalam memastikan konsistensi dan akurasi data analisis di industri farmasi. Laporan ini membahas pengembangan protokol Kualifikasi Instalasi (KI) dan Kualifikasi Operasional (KO) untuk alat Micro Balance di PT. Pfizer Indonesia, yang direlokasi pada Desember 2023. Relokasi alat tersebut memerlukan kualifikasi untuk menjamin alat tetap berfungsi sesuai spesifikasi yang ditetapkan. Pendekatan metodologi menggunakan Failure Modes, Effects, and Criticality Analysis (FMECA) untuk menilai risiko terkait kompleksitas alat, sensitivitas, dan rekomendasi produsen. Berdasarkan tinjauan risiko, Micro Balance diklasifikasikan sebagai medium risk, sehingga memerlukan kualifikasi instalasi dan operasional. Protokol yang disusun mencakup verifikasi parameter teknis seperti identifikasi alat, stabilitas lingkungan, serta pengujian operasional yang melibatkan pengukuran berat terkalibrasi. Proses ini memastikan alat beroperasi optimal sesuai regulasi Good Manufacturing Practices (cGMP). Kesimpulan menunjukkan bahwa pelaksanaan kualifikasi secara sistematis mampu menjaga mutu alat selama dan setelah relokasi. Protokol ini dapat menjadi acuan dalam implementasi serupa di masa depan, mendukung kepatuhan terhadap standar Cara Pembuatan Obat yang Baik (CPOB) yang ditetapkan oleh BPOM.

Equipment qualification is a crucial step in ensuring the consistency, and accuracy of analytical data in the pharmaceutical industry. This report discusses the development of Installation Qualification (IQ) and Operational Qualification (OQ) protocols for a Micro Balance device at PT. Pfizer Indonesia, which was relocated in December 2023. The relocation required qualification to ensure the device operates according to the specified standards. The methodology employed Failure Modes, Effects, and Criticality Analysis (FMECA) to assess risks related to the device's complexity, sensitivity, and manufacturer recommendations. Based on the risk assessment, the Micro Balance was classified as medium risk, necessitating installation and operational qualifications. The protocols developed included verification of technical parameters such as device identification, environmental stability, and operational testing involving calibrated weight measurements. This process ensures the device functions optimally under Good Manufacturing Practices (cGMP) regulations. The findings conclude that systematic qualification implementation maintains equipment quality during and after relocation. This protocol can serve as a reference for similar implementations in the future, supporting compliance with Good Manufacturing Practices (GMP) standards established by the Indonesian Food and Drug Authority (BPOM).

 File Digital: 1

Shelf
 PR-Muhammad Mishbahus Surur.pdf :: Unduh

LOGIN required

 Metadata

Jenis Koleksi : UI - Tugas Akhir
No. Panggil : PR-pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Farmasi Universitas Indonesia, 2024
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : vii, 18 pages : illustration + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
PR-pdf 16-25-44558539 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 9999920566953
Cover