Evaluasi keluaran sinyal generator tabung sinar-x menggunakan metode non-invasive = Evaluation of the x-ray tube output signals generators by non invasive methods

Cahyo Fitriyani Maisyaroh; Prawito Prajitno, supervisor; Heru Prasetio, supervisor; Djarwani Soeharso Soejoko, examiner; Dwi Seno Kuncoro Sihono, examiner (Universitas Indonesia, 2012)

 Abstrak

ABSTRAK
Generator sinar-x menyuplai energi untuk tabung sinar-x, yang berfungsi untuk
mengatur parameter paparan radiasi. Dilakukan evaluasi menggunakan metode
non-invasive untuk mendeteksi jenis ketidaksesuaian fungsi sistem loading,
rectifier, contactor, shutter, cable ataupun connector arcing pada generator dari
bentuk sinyal yang dihasilkan pada program Xi View Unfors dan dilakukan juga
evaluasi terhadap kinerja generator. Penilaian hasil uji fungsi ini dilakukan
dengan melakukan analisa fungsi generator sinar-x menggunakan data yang telah
diukur di PTKMR meliputi berbagai jenis generator constant potential dan singlephase.
Detektor unfors diletakkan pada jarak 100 cm dari fokus sumber radiasi
pesawat sinar-x. Tegangan panel pesawat bervariasi mulai dari kV terendah
sampai kV tertinggi. Parameter uji yang dilakukan dalam uji fungsi pesawat
sinar-x pada penelitian ini hanya yang berkaitan dengan kondisi penyinaran yaitu
akurasi tegangan kerja (kV), akurasi waktu paparan (s), kesesuaian nilai HVL
(Half Value Layer). Hasil dari uji fungsi dibandingkan dengan referensi dari
Western Australia. Berdasarkan evaluasi yang dilakukan, pesawat sinar-x yang
lulus uji fungsi, tidak menjamin memiliki fungsi generator yang sesuai dengan
standar spesifikasi. Kemungkinan surge step, overshoot dan contactor bounce
kerap terjadi pada awal paparan, sedangkan drop tegangan maupun discharge
kerap terjadi pada saat paparan berlangsung.

ABSTRACT
The x-ray generator suplies energy to the x-ray tube, that controlling the radiation
exposure parameter. Evaluation done using non-invasive methods to detect
generators problems at system function such as loading, rectifier, contactor,
shutter, cable and connector arcing from signal output showed on Xi View Unfors
program. We also evaluate the generator performances. Compliance test done by
analyzing the x-ray generator function that measured by PTKMR, which consist
of constant potential generator and single-phase generator. Unfors detector placed
at 100 cm away from the focus of x-ray radiation source. Variation of panel
voltage start from the lower kV to Higher kV. Compliance test parameters in this
evaluation only related with exposure influence parameter, which are panel
voltage accuracy (kV), exposure times accuracy (s), and HVL. This results will
be compare to Western Australia?s standard. According to the evaluation, x-ray
machines that passed the compliance test did not guaranteed that the generator has
performance as specified by specification. Surge step, overshoot and contactor
bounce is happened at the first of the exposure and votage droop, discharges is
happened when the exposure running.

 File Digital: 1

 Metadata

No. Panggil : S42145
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Subjek :
Penerbitan : [Place of publication not identified]: Universitas Indonesia, 2012
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resources
Deskripsi Fisik : xii, 49 pages : illustration ; 30 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
S42145 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20304074