Karakterisasi detektor in vivo untuk dosimetri radioterapi eksterna = Characterisation of In vivo detector for external radiotherapy dosimetry

Aden Rendang Sumedi Putri; Dwi Seno Kuncoro Sihono, supervisor; Heru Prasetio, supervisor; Djarwani Soeharso Soejoko, examiner; Sugiyntari, examiner (Universitas Indonesia, 2012)

 Abstrak

ABSTRAK
Sistem dosimetri in vivo dengan menggunakan detektor dioda telah dikaraterisasi
untuk estimasi dosis pasien. Penelitian ini dilakukan untuk menentukan
karakteristik detektor dioda dengan menentukan faktor kalibrasi dioda dan faktor
koreksi dengan beberapa variasi. Langkah penelitian ini antara lain menentukan
kesetaraan virtual water terhadap air, faktor kalibrasi dioda, linearitas dosis,
koreksi luas lapangan, koreksi SSD, koreksi sudut dan koreksi tray. Pengukuran
karakterisasi dioda menghasilkan Faktor kalibrasi dioda pada energi 6 MV sebesar
3,189 cGy/nC dan pada 10 MV sebesar 3,121 cGy/nC. Detektor dioda pada dosis
10-300 cGy memiliki respon linear terhadap perubahan dosis. Variasi luas
lapangan memiliki presentase kurang dari 2 %, kecuali pada energi 10 MV
dengan luas lapangan 4 x 4 cm2, 35 x 35 dan 40 x 40 cm2. Pada energi 6 MV dan
10 MV semakin pendek jarak, faktor koreksi SSD semakin kecil. Arah datang
sinar dari sudut -50 s/d 50 derajat pada energi 6 MV dan 10 MV memberikan efek
yang relatif konstan,dan efek arah datang sinar meningkat tajam dimulai dari
sudut 60 derajat. Koreksi tray untuk setiap luas lapangan kecuali luas lapangan 4 x
4 cm2 relatif konstan sehingga cukup koreksi dengan satu kondisi saja.

ABSTRACT
An in vivo dosimetry system that used diode detector was characterized for dose
estimates patient. The measurenment was done to determine diode detector
characterisation with find calibration factor and correction factor with several
variation. Step of this research among others determine virtual water equality to
water, diode calibration factor, dose linearity, field size correction, SSD, angular,
and tray correction. Diode characterisation measurantment produce 3,189 cGy/nC
diode calibration factor at 6 MV and 3,121 cGy/nC at 10 MV. Diode detector has
linearity responds at 10 -300 cGy dose. correction factor of field size less to 2 %
except field size of 4 x 4, 35 x 35, and 40 x 40 cm2. shorter SSD has less
correction factor , the angular effect at -50 up to 50 degrees is constant relatively,
and the effect increases start at 60 degrees. Using single tray condition to
determine tray correction factor

 File Digital: 1

Shelf
 S42499-Karakterisasi detektor.pdf :: Unduh

 Metadata

No. Panggil : S42499
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Penerbitan : [Place of publication not identified]: Universitas Indonesia, 2012
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resources
Deskripsi Fisik : xii, 41 pages : illustration ; 29 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
S42499 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20308590