Pengukuran dosis organ kritis gonad dan tyroid serta ESD pada prosedur fluoroskopi konvensional = Measurement critical organ dose gonad and thyroid and ESD on fluoroscopy conventional procedure / Edhy Kuntowibowo

Edhy Kuntowibowo; Djarwani Soeharso Soejoko, promotor; Mussadiq Musbach, examiner; Arman Adel Abdullah, examiner (Universitas Indonesia, 2013)

 Abstrak

ABSTRAK
Fluoroskopi adalah pencitraan real time yang dapat memperoleh informasi organ
tubuh yang bergerak dan memiliki resiko radiasi relatif lebih tinggi dari radiografi
pada umumnya. Telah dilakukan pengukuran dosis pada beberapa prosedur
fluoroskopi konvensional di organ kritis tyroid dan gonad serta ESD
menggunakan TLD dan jumlah produksi sinar-X dengan DAP meter. Jumlah
sampel 36 orang pasien terdiri dari 16 sampel HSG, 10 apendiks, 6 fistulografi,
dan 4 Colon In Loop . Pengukuran dosis organ kritis dilakukan dengan
meletakkan TLD pada permukaan kulit pasien di 3 lokasi yaitu daerah tyroid,
gonad, dan ESD. Hasil pengukuran menunjukkan dosis tyroid menerima dosis
rendah dengan rata-rata sedikit di atas orde latar. Adapun rata-rata dosis gonad
paling tinggi diperoleh diperoleh dari prosedur fistulografi yaitu 5,58 ± 4,64 mGy
dan dalam rentang 1,36 – 11, 66 mGy sedangkan prosedur yang lain lebih rendah.
Selanjutnya diperoleh nilai DAP yang bervariasi akibat ketebalan obyek yang
tidak konstan dan waktu fluoroskopi yang bervariasi, nilai DAP dari 16 sampel
pasien HSG dalam rentang 0,2 – 0,7 mGym2, 10 sampel apendiks dalam rentang
0,1 – 0,64 mGym2, 6 sampel fistulografi dalam rentang 0,14 – 1,33 mGym2, dan 4
sampel Colon In Loop dalam rentang 0,86 – 3,44 mGym2.

ABSTRACT
Fluoroscopy is an real time imaging which can obtain information organs are
moving and have a relatively higher risk of radiation in general radiography.
Dose measurements have been carried out on some conventional fluoroscopy
procedures in critical organ for thyroid and gonads and ESD using TLD and
quantity production of X-rays with a DAP meter. Number of samples 36 patients
consisted of 16 samples HSG, 10 appendices, 6 fistulografi, and 4 collon in the
loop. Critical organ dose measurements performed with TLD on the surface of
skin putting patients at 3 locations: thyroid, gonads, and ESD. The measurement
results indicate thyroid doses received low dose with an average slightly above
background order. The average of the highest gonad doses obtained fistulografi
procedure is obtained 5.58 ± 4.64 mGy and in the range 1.36 - 11, 66 mGy while
the other procedures is lower. Furthermore DAP values obtained varied due to
the thickness of the object is not constant and varied fluoroscopy time, DAP value
of 16 patient samples HSG in the range 0.2 - 0.7 mGym2, appendix 10 samples in
the range of 0.1 - 0.64 mGym2, fistulografi 6 samples in the range of 0.14 - 1.33
mGym2, and 4 samples collon in the loop in the range of 0.86 - 3.44 mGym2.

 File Digital: 1

Shelf
 T35623-Edhy Kuntowibowo.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : T35623
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Subjek :
Penerbitan : [Place of publication not identified]: Universitas Indonesia, 2013
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xiii, 38 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
T35623 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20349796