Verifikasi geometri radioterapi teknik 3DCRT/IMRT pada kasus kanker kepala dan leher di Departemen Radioterapi RSCM = Geometric verification of head and neck cancer radiotherapy using 3DCRT/IMRT in Department of Radiotherapy, Cipto Mangunkusumo Hospital

(Fakultas Kedokteran Universitas Indonesia, 2013)

 Abstrak

[Pendahuluan : Radioterapi pada kanker kepala dan leher menggunakan teknik
Three-dimensional Conformal Radiotherapy (3DCRT) atau Intensity-modulated
Radiotherapy (IMRT) membutuhkan akurasi yang tinggi dalam pelaksanaannya.
Upaya ini dilakukan dengan mengetahui kesalahan set-up melalui proses
verifikasi yang disesuaikan dengan beban kerja setiap unit radioterapi. Dengan
demikian dapat diterapkan margin CTV-ke-PTV yang ideal untuk mendapatkan
dosis yang adekuat pada area target radiasi.
Metode penelitian : Penelitian ini merupakan studi potong lintang yang
mengambil data verifikasi menggunakan Cone Beam Computed Tomography
(CBCT) dari 9 pasien kanker kepala dan leher yang mendapatkan radioterapi
dengan teknik 3DCRT/IMRT di Departemen Radioterapi Rumah Sakit Cipto
Mangunkusumo (RSCM) antara bulan Oktober 2013 hingga Desember 2013.
Pergeseran pada lapangan radiasi yang didapatkan dari hasil verifikasi dalam lima
fraksi awal dianalisis untuk memperoleh kesalahan sistematik dan kesalahan acak,
yang selanjutnya dihitung untuk mendapatkan margin CTV-ke-PTV.
Hasil : Sebanyak 135 data verifikasi CBCT dianalisa. Besar kesalahan sistematik
dan kesalahan acak yang didapatkan berturut-turut sebesar 1.5 mm dan 2.7 mm
pada sumbu laterolateral, 2.2 mm dan 3.1 mm pada sumbu kraniokaudal, serta 2.2
mm dan 1.9 mm untuk sumbu anteroposterior. Margin CTV-ke-PTV yang
diperoleh sebesar 4.9 mm, 6.6 mm dan 5.8 mm untuk masing-masing sumbu
laterolateral, kraniokaudal dan anteroposterior.
Kesimpulan : Verifikasi menggunakan CBCT dalam lima fraksi awal merupakan
metode yang efektif untuk deteksi dan koreksi kesalahan set-up. Hasil penelitian
dapat digunakan sebagai rekomendasi pemberian margin CTV-ke-PTV dan
menunjukkan pemberian margin sebesar 5 mm sudah cukup adekuat dalam
pelaksanaan radioterapi kanker kepala dan leher dengan teknik 3DCRT/IMRT di
Departemen Radioterapi RSCM. Diperlukan upaya tambahan untuk meningkatkan
koreksi kesalahan set-up dengan memperhitungkan beban kerja unit radioterapi.
Kata kunci : kanker kepala dan leher, radioterapi teknik 3DCRT/IMRT,, Introduction : Three-dimensional Conformal Radiotherapy (3DCRT) or Intensity
Modulated Radiotherapy (IMRT) for head and neck cancer is a highly accurate
procedure. Verification is needed to detect and correct set-up errors, adjusted
according to workload of each radiotherapy center. Therefore, an ideal CTV-to-
PTV margin can be applied to ensure adequate target volume coverage.
Methods : This is a cross-sectional study using Cone Beam Computed
Tomography (CBCT) verification data of 9 head and neck cancer patients treated
with 3DCRT/IMRT in Department of Radiotherapy, Cipto Mangunkusumo
Hospital between October 2013 and December 2013. Translation errors from the
first five fractions were analyzed to count for systematic and random errors.
These errors were then calculated to acquire CTV-to-PTV margin.
Results : A total of 135 CBCT data were analyzed. Systematic and random errors
were respectively 1.5 mm and 2.7 mm in laterolateral direction, 2.2 mm and 3.1
mm in craniocaudal direction, and 2.2 mm and 1.9 mm in anteroposterior
direction. CTV-to-PTV margin were 4.9 mm, 6.6 mm and 5.8 mm in laterolateral,
craniocaudal and anteroposterior direction, respectively.
Conclusions : CBCT verification in first five fractions was effective in detecting
and correcting set-up errors. The calculated CTV-to-PTV margin can be used as
recommended margin and showed that 5 mm margin was adequate in planning
3DCRT/IMRT technique for head and neck cancer in Department of
Radiotherapy, Cipto Mangunkusumo Hospital. An extra effort might be done to
improve the correction of set-up errors adjusted to workload.]

 File Digital: 1

Shelf
 SP-Faisal Adam.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : SP-pdf
Subjek :
Penerbitan : [Place of publication not identified]: Fakultas Kedokteran Universitas Indonesia, 2013
Program Studi :
Bahasa : ind
Sumber Pengatalogan :
Tipe Konten :
Tipe Media :
Tipe Carrier :
Deskripsi Fisik : xii, 30 hlm. : ill. ; 28 cm. + lamp.
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
SP-pdf TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20367229