:: UI - Skripsi Membership :: Kembali

UI - Skripsi Membership :: Kembali

Validasi metode analisis orto kresol dan penentuan kadarnya dalam urin dengan menggunakan instrumen kromatografi gas = Ortho cresol validation in urine using gas chromatography instrument / Lina Setiawati

Lina Setiawati; Sunardi, supervisor; Antonius Herry Cahyana, examiner; Rahmat Wibowo, examiner; Budiawan, examiner (Universitas Indonesia, 2014)

 Abstrak

ABSTRAK
Sebuah metode GC telah divalidasi dan akan digunakan untuk penentuan kadar o-kresol yang terkandung dalam sampel urin para perokok pasif dan perokok aktif sebagai biomarker paparan toluen. Sampel yang akan dianalisis diekstraksi dahulu untuk kemudian disuntikan ke GC. Metode ini memenuhi kriteria validasi dilihat dari linearitas, rentang, presisi, batas deteksi dan persen perolehan kembali. Diperoleh nilai koefisien regresi sebesar 0,996. Nilai akurasi dinyatakan dengan persen perolehan kembali, di dapat nilai sebesar 93,2%. Untuk uji presisi dilakukan metode keterulangan, dinyatakan dengan %RSD yang memberikan nilai sebesar 5,76%, dianggap telah memenuhi standar yaitu %RSD ≤ 16%. Diperoleh nilai batas deteksi 2,98ppm. Pencapaian penelitian validasi instrumen GC ini memberikan hasil yang cukup memuaskan, maka metode ini cocok untuk digunakan sebagai pemantau kadar o-kesol dalam sampel urin untuk meninjau paparan toluen. Metode ini diterapkan pada tujuh sukarelawan, terdiri dari 3 perokok pasif dan 4 perokok aktif. Hasil yang didapatkan setelah semua sempel di uji memberikan nilai kadar o-kresol pada rentang konsentrasi 13,04ppm sampai 39,40ppm.

ABSTRACT
A validation of gas chromatography method was performed for determination of ortho-cresol contained in urine of passive smokers and active smokers as biomarkers of toluene exposure . After extraction , the samples were analyzed by GC instrumen. This method has met the validation criteria by the terms of linearity, range, precision, limit of detection and percent recovery. Regression coefficient values was obtain for 0,996. Accuracy value are expressed as percent recovery is able to reach 93.2 %. For precision test is using repeatibility method, expressed by %RSD and give a value 5.76. In limit of detection value was owned for 2.98 ppm. Achievement of a validation gas chromatography instrument in this study has provided satisfactory results, so this method is considered suitable for use as a monitors levels of ortho-cresol in urine to review toluene exposure . This method is applied to seven volunteers consisting of three of passive smokers and four of active smokers. The results in this thesis obtained an ortho-cresol levels inside the sample is between 13,04ppm until 39,40ppm

 File Digital: 1

Shelf
 S53233-Lina Setiawati.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : S53233
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Subjek :
Penerbitan : [Place of publication not identified]: Universitas Indonesia, 2014
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xv, 53 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
S53233 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20368096