Geometric accuracy of stereotactic radiotherapy in Department of Radiotherapy in Cipto Mangunkusumo Hospital = Geometric accuracy of stereotactic radiotherapy in Department of Radiotherapy in Cipto Mangunkusumo Hospital

(Fakultas Kedokteran Universitas Indonesia, 2014)

 Abstrak

[Pendahuluan : Radioterapi stereotaktik adalah suatu bentuk terapi radiasi yang
membutuhkan akurasi tinggi. Selain imobilisasi yang baik, dibutuhkan verifikasi untuk
memastikan akurasi dan untuk mengetahui kesalahan sistematik dan acak dalam pemberian
radiasi. Margin Planning Target Volume (PTV) dibuat untuk memastikan target radiasi
mendapatkan cakupan dosis radiasi yang diinginkan.
Metode penelitian : Penelitian ini merupakan studi retrospektif yang menggunakan data
verifikasi dengan X-ray Volumetric Imaging (XVI) dari 10 pasien yang menjalani radioterapi
stereotaktik di Departemen Radioterapi Rumah Sakit Cipto Mangunkusumo (RSCM) dengan
fiksasi bite-block antara bulan Januari 2013 sampai dengan Oktober 2013. Penyimpangan
yang didapat digunakan untuk menghitung kesalahan sistematik dan kesalahan acak. Margin
PTV didapat dengan memasukkan kesalahan sistematik dan kesalahan acak yang didapat ke
dalam formulasi Stroom.
Hasil : Sebanyak 42 hasil XVI dianalisa. Pada penelitian ini didapatkan besar kesalahan
sistematik dan kesalahan acak pada pasien yang menjalani radioterapi stereotaktik di RSCM
adalah sebesar 0.61 mm dan 1.27 mm untuk sumbu laterolateral, 1.13 mm dan 2.41 mm
untuk sumbu kraniokaudal, serta 0.71 mm dan 0.15 mm untuk sumbu anteroposterior. Margin
PTV yang diperoleh sebesar 2.11 mm, 3.95 mm dan 2.22 mm untuk masing-masing sumbu
laterolateral, kraniokaudal dan anteroposterior.
Kesimpulan : Hasil penelitian memberikan rekomendasi margin PTV yang dapat digunakan
di Departemen Radioterapi RSCM. Terdapat margin PTV yang cukup besar untuk sumbu
kraniokaudal. Diperlukan upaya tambahan untuk meningkatkan akurasi radiasi sehingga
margin yang diberikan tidak terlalu besar., Introduction : Stereotactic radiotherapy is a technique to administer precisely directed
irradiation that tightly conforms to a target volume. Beside immobilisation, verfication is
needed to ensure the accuracy and to calculate the systematic and random error. Planning
Target Volume (PTV) margin is delineated to ensure adequate target volume coverage.
Methods : This is a retrospective study using X-ray Volumetric Imaging (XVI) data of 10
patients who have had stereotactic radiotherapy with bite-block fixation between January
2013 and October 2013 in the Department of Radiotherapy in Cipto Mangunkusumo
Hospital. The translation errors were used to calculate the systematic and random error.
PTV margin was acquired by using these errors in Stroom formula.
Results : A total of 42 XVI were analyzed. Systematic and random errors were respectively
0.61 mm and 1.27 mm in laterolateral direction, 1.13 mm and 2.41 mm in craniocaudal
direction, and 0.71 mm and 0.15 mm in anteroposterior direction. PTV margin were 2.11
mm, 3.95 mm and 2.22 mm in laterolateral, craniocaudal and anteroposterior direction
respectively.
Conclusions : The result gave PTV margin recommendation that can be used in Department
of Radiotherapy in Cipto Mangunkusumo Hospital. There was a quite large PTV margin in
craniocaudal direction. Some efforts and evaluations are needed to improve the accuracy of
stereotactic radiotherapy to reduce the PTV margin.]

 File Digital: 1

Shelf
 SP-Afred Julius.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : SP-pdf
Penerbitan : [Place of publication not identified]: Fakultas Kedokteran Universitas Indonesia, 2014
Program Studi :
Bahasa : ind
Sumber Pengatalogan :
Tipe Konten :
Tipe Media :
Tipe Carrier :
Deskripsi Fisik : xii, 30 hlm. : ill. ; 28 cm. + lamp.
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
SP-pdf TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20377761