Full Description

Cataloguing Source :
ISSN : 20891814
Magazine/Journal : Jurnal eksplora informatika 3 (1) September 2013. Hal. : 39-48
Volume :
Content Type :
Media Type :
Carrier Type :
Electronic Access :
Holding Company : Universitas Indonesia
Location : Perpustakaan UI, Lantai 4, R. Koleksi Jurnal
 
  •  Availability
  •  Digital Files: 0
  •  Review
  •  Cover
  •  Abstract
Call Number Barcode Number Availability
005 JEI 3:1 (2013) TERSEDIA
No review available for this collection: 20439900
 Abstract
The research was concerned with creating a device for vocal signal monitoring based on microcontroller application. The device was implemented to monitor the vocal signal in order to prevent a negative impact so called voice disorder. The availability of a device for vocal signal monitoring can be a benefit to the concerned users. The principal work was creating a suitable logger system, including software and hardware parts. The hardware part consists of logger system, conditioning circuitry, and power circuitry. The data obtained from the microcontroller was analyzed off-line using MATLAB. The device was organized to work in both normal and calibration modes. The analysis consisted of Sound Pressure Level (SPL) calibration and fundamental frequency (f0) estimation which are the most important parameters used in voice monitoring systems. According to the experiment, the calibration constant K was 1.76 Pa/V (0.12% standard deviation) and error average of the ECM was -1.8 dB (4.6 dB standard deviation) compared to the reference microphone.Collaboration works between engineers and physicians are advisable for a correct use of the estimated parameters.