Penggunaan ultrasonografi untuk konfirmasi tip peripheraly inserted central catheter dibandingkan dengan radiografi standar = The use of ultrasound to confirm tip of peripherally inserted central catheters compared to standard radiograph / James Thimoty

James Thimoty; Evita Karijani Bermanshah, supervisor; Rinawati Rohsiswatmo, supervisor; Pardede, Sudung O., examiner; Mulyadi M. Djer, examiner; Rosalina Dewi Roeslaini, examiner; Toto Wisnu Hendarto, examiner ([Publisher not identified] , 2017)

 Abstrak

ABSTRAK
Latar Belakang. Penggunaan peripherally inserted central catheter PICC semakin meningkat sesuai kebutuhan bayi prematur dalam pemberian nutrisi parenteral. Penggunaan radiografi standar Rontgen torako abdomen dalam menilai posisi tip PICC yang tepat merupakan baku emas. Namun demikian, radiografi standar tidak selalu akurat dalam memperkirakan posisi tip PICC, tergantung pada posisi ekstremitas dan variasi anatomi masing-masing neonatus, menyebabkan radiasi, membutuhkan biaya yang lebih tinggi, waktu yang lebih lama dan membutuhkan pemeriksaan ulangan apabila posisi tip PICC terlalu dalam masuk ke jantung. Ultrasonografi USG semakin banyak digunakan untuk diagnostik dan prosedur di neonatal intensive care unit NICU karena lebih mudah, aman, murah dan real-time.Tujuan. Penelitian ini bertujuan untuk menilai akurasi penggunaan USG dalam menentukan ketepatan posisi tip PICC dibandingkan dengan radiografi standar. Metode. Penelitian ini merupakan studi diagnostik yang dilakukan pada bulan April-Mei 2017 di NICU. Pemasangan PICC menggunakan prosedur standar di NICU dan kemudian posisi tip PICC dievaluasi menggunakan USG sebelum dilakukan radiografi standar. Analisis menggunakan table 2x2 untuk menilai akurasi diagnostik USG dibandingkan radiografi standar dalam menilai posisi tip PICC.Hasil. Sebanyak 29 neonatus termasuk dalam penelitian ini dengan rerata usia gestasi 31,7 minggu 26 sampai 41 minggu dan rerata berat lahir 1618,9 g 600 sampai 3750 g . Kesesuaian antara radiografi standar dan ultrasonografi untuk posisi tip PICC terjadi pada 27 neonatus 93,1 . Sensitivitas ultrasonografi dalam menentukan tip PICC adalah 88,89 , dengan spesifisitas 95 dan ultrasonografi memiliki akurasi diagnostik 93,1 .Simpulan. USG memiliki akurasi diagnostik yang baik untuk mengindentifikasi posisi tip PICC, meminimalkan paparan radiasi dan meminimalkan manipulasi PICC di NICU.Kata kunci: USG, Radiografi, PICC, Neonatus, NICU

ABSTRACT
Background. The use of peripherally inserted central catheter PICC has became increasingly common to facilitate the administration of parenteral nutrition in preterm neonates. The use of standard radiograph X ray of the abdominal thorax in assessing the correct position of peripherally inserted central catheter PICC line tips is a gold standard. However, standard radiograph can not always be accurate depending on the position of the extremities and variations of anatomy of each neonates, cause radiation, require higher cost, longer time and repeat when position is not appropriate. Ultrasound are becoming increasingly use for diagnostic and procedure in neonatal intensive care unit NICU because more easy, safer, cheaper, and real time.Objectives. This study aims to assess the accuracy of ultrasound use in determining the accuracy of PICC tip positions compared to standard radiograph.Methods. This was a diagnostic study to examine the placement of PICC in neonates between April and May 2017 in NICU. PICC were placed using standard procedure in NICU and then position of PICC tip were evaluated using ultrasound immediately before standard radiograph is performed. A 2x2 table was constructed to compare the two modalities ability to detect tip PICC position.Results. A total of 29 neonates were included in this study with mean of gestation of 31.7 weeks 26 to 41 weeks and mean of birth weight 1618.9 g 600 to 3750 g . Concordance between standard radiograph and ultrasound for PICC tip position occurred in 27 neonates 93.1 . Sensitivity of ultrasound in determining PICC tip was 88.89 , with a specificity of 95 and ultrasound had a diagnostic accuracy 93.1 .Conclusion. Ultrasound has a good diagnostic accuracy to indentify tip position of PICC, minimizing exposure of radiation and minimize manipulation of PICC in NICU. Keywords Ultrasound, Radiograph, PICC, Neonates, NICU

 File Digital: 1

Shelf
 T-James Thimoty.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : T-Pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Subjek :
Penerbitan : [Place of publication not identified]: [Publisher not identified], 2017
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : xv, 83 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
T-Pdf 15-20-485478565 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20455985