Uji kesesuaian pengukuran oksimeter denyut jari dengan modifikasi oksimeter denyut orofaring = Agreement test of measurement between finger pulse oxymeter and modification oropharyngeal pulse oxymeter

Andreas Robby Susanto; Aldy Heriwardito, supervisor; Amir Sjarifuddin Madjid, supervisor; Ratna Farida Soenarto, examiner; Indro Mulyono, examiner; Aries Perdana, examiner; Dita Aditianingsih, examiner (Universitas Indonesia, 2017)

 Abstrak

Latar Belakang. Saturasi oksigen merupakan pengukuran rutin pada pasien dengan pipa endotrakea ataupun dengan masalah pernafasan / oksigenasi dan merupakan standar pemantauan oksigenasi dari ASA dan WHO. Oksimeter denyut jari menjadi pilihan utama karena mudah digunakan. Pemantauan saturasi oksigen sering bermasalah karena lokasi di perifer terdapat kelainan , misalnya karena luka bakar, hipoperfusi. Diperlukan alternatif pemantauan saturasi antara lain dengan modifikasi oksimeter denyut orofaring. Penelitian ini bertujuan untuk membandingkan kesesuaian hasil pengukuran antara modifikasi oksimeter denyut orofaring dan oksimeter denyut jari. Metode. Penelitian analitik observasional dengan rancangan potong lintang terhadap pasien ASA 1 dan ASA 2 terintubasi yang menjalani pembedahan di RSCM Jakarta periode Agustus-September 2017. Sebanyak 26 pasien diambil secara konsekutif, setiap pasien dilakukan dua pengukuran saturasi dengan oksimeter denyut jari dan modifikasi oksimeter denyut orofaring, pencatatan dilakukan setiap 5 menit selama 30 menit . Analisis data menggunakan uji kesesuain Bland-Altman. Hasil. Rerata hasil pengukuran oksimeter jari sebesar 98,53 SD 0,896 , dan rerata hasil pengukuran modifikasi oksimeter denyut orofaring sebesar 98,35 SD 1,238 . Analisa uji Bland Altman pada semua waktu pengukuran mendapatkan rerata selisih antara oksimeter denyut jari dan modifikasi oksimeter denyut orofaring sebesar 0,19 , dengan interval kepercayaan 95 sebesar 0.07-1,79 dan dengan limit of agreement sebesar -1,42 ndash; 1,79. Simpulan. Pengukuran dengan metode modifikasi oksimeter denyut orofaring memiliki kesesuaian yang sangat baik dengan oksimeter denyut jari.

Background. Oxygen saturation is an important routine measurement in patients with endotracheal tubes or with respiratory oxygenation problems and standard oxygenation monitoring from ASA and WHO. Finger pulse oxymeter is the first choice because of its easy use. Oxygen saturation monitoring often has problems if the peripheral site has abnormalities due to several things, because. Alternative saturation monitoring is required, modification oropharyngeal pulse oxymeter is one choice. This study aims to compare the agreement of the measurement results between the modification of the oropharynx pulse oximeter and the finger pulse oximeter. Method. An observational analytic study with cross sectional design which observe ASA 1 and ASA 2 patients who were intubated and undergo surgery at RSCM Jakarta period August September 2017. Total 26 patients were taken consecutively, each patient were performed finger pulse oximeter and modification oropharyngeal pulse oxymeter, recorded every 5 minutes for 30 minutes. Analysis using agreement test Bland Altman. Results. The mean of measurement with finger oximeter was 98,53 SD 0,896 , and mean of measurement result with modification oropharyngeal pulse oximeter was 98,35 SD 1,238 . Analysis with Bland Altman test at all measurement time got mean difference between finger pulse oximeter and modification oropharyngeal pulse oximeter of 0.19, with 95 confidence interval equal to 0.07 1,79 and with limit of agreement equal to 1,42 1,79. Conclusion. Measurements with the modification oropharyngeal pulse oximeter method have good agreement with finger pulse oximeter.

 File Digital: 1

Shelf
 T-Andreas Robby Susanto.pdf :: Unduh

LOGIN required

 Metadata

No. Panggil : T-Pdf
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Entri tambahan-Nama badan :
Subjek :
Penerbitan : Depok: Universitas Indonesia, 2017
Program Studi :
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : computer
Tipe Carrier : online resource
Deskripsi Fisik : xiv, 68 pages : illustration ; 28 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
No. Panggil No. Barkod Ketersediaan
T-Pdf 15-18-653062210 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20467508