Perbandingan nilai amplitudo dan latensi antara Elektroda Jet, Elektroda Dencott dan Elektroda Dawson-Trick-Litzkow pada pemeriksaan elektroretinografi multifokal mata normal pada populasi orang dewasa di Rumah Sakit Umum Pusat Nasional Cipto Mangunkusumo = Comparison of amplitude and implicit time between Jet Electrode, Dencott Electrode and Dawson-Trick-Litzkow Electrode in multifocal electroretinography examination in adult in Cipto Mangunkusumo National hospital
Ika Citra Susanti;
M. Sidik, supervisor; Syntia Nusanti, supervisor; Aria Kekalih, supervisor; Andi Arus Victor, examiner; Gusti Gede Suardana, examiner; Yunia Irawati, examiner; Neni Anggraini, examiner; Anna Puspitasari Bani, examiner
([Publisher not identified]
, 2018)
|
ABSTRAK Mendapatkan nilai standar normal pemeriksaan elektroretinografi multifokal danmencari perbandingan nilai tersebut berikut tingkat kenyamanan pasien menggunakanelektroda Jet, Dencott dan Dawson-Trick-Litzkow pada populasi orang Indonesiadewasa di RSUPNCM. Melalui metode convenient sampling dari 49 subjek orangIndonesia berusia 19-49 tahun dengan kedua mata normal, dilakukan pengukuranamplitudo dan latensi gelombang berdasarkan rekomendasi ISCEV. Pengukuranmeliputi gelombang N1 dan P1 pada ring 1 sampai 5. Terdapat perbedaan signifikannilai standar normal pemeriksaan elektroretinografi multifokal pada populasi orangIndonesia dewasa secara statistik antara elektroda dengan amplitudo gelombang lebihtinggi dan latensi lebih panjang pada elektroda Jet dan Dencott dibandingkan denganDTL. Elektroda DTL dan Jet dirasakan lebih nyaman daripada elektroda Dencott bagiorang Indonesia dewasa normal untuk pemeriksaan elektroretinografi multifokal.Elektroda DTL memberikan amplitudo gelombang yang paling rendah, latensi yangapling pendek dan tingkat kenyamanan yang paling tinggi dibandingkan elektroda Jetdan Dencott dalam pemeriksaan elektroretinografi multifokal pada orang Indonesiadewasa normal. ABSTRACT To establish normal values of multifocal electroretinography (ERG) and to comparethe values and the comfort level using Jet, Dencott and DTL electrode in IndonesianAdult. Through convenient sampling 49 normal Indonesian subjects between 19 and49 yeras oldwere selected. Multifocal ERG amplitudes and implicit time values weremeasured according to recommendation by the International Society for ClinicalElectrophysiology of Vision. Evaluation consisted of N1 and P1 wave in ring 1 to 5.after the examination, all subjects filled in a questionnaire about comfort level,adopted from the visual analog scale. We observed a statistically significantdifference in multifocal ERG normal values between electrode with higher waveamplitudes and longer implicit time in Jet and Dencott electrodes compared to DTLelectrodes. Jet and DTL electrodes are more comfortable than Dencott electrodes forIndonesian adults in multifocal ERG. DTL electrodes give the lowest wave amplitudeand the shortest implicit time and are the most comfortable electrode compared to Jetand Dencott electrodes, in multifocal ERG in Indonesian adults. |
![]()
|
No. Panggil : | SP-pdf |
Entri utama-Nama orang : | |
Entri tambahan-Nama orang : | |
Entri tambahan-Nama badan : | |
Subjek : | |
Penerbitan : | [Place of publication not identified]: [Publisher not identified], 2018 |
Program Studi : |
Bahasa : | ind |
Sumber Pengatalogan : | |
Tipe Konten : | text |
Tipe Media : | computer |
Tipe Carrier : | online resource |
Deskripsi Fisik : | xiii, 61 pages : illustration ; 28 cm + appendix |
Naskah Ringkas : | |
Lembaga Pemilik : | Universitas Indonesia |
Lokasi : | Perpustakaan UI, Lantai 3 |
No. Panggil | No. Barkod | Ketersediaan |
---|---|---|
SP-pdf | 16-21-562883288 | TERSEDIA |
Ulasan: |
Tidak ada ulasan pada koleksi ini: 20485509 |